SPECIFICATIONS
OF HYDROGEN FLUORIDE TWO WIRE 4-20 mA TRANSMITTER
UNIPHOS
– 572 T
Gas
:
Hydrogen Fluoride (HF)
Range
(ppm)
:
0 – 10 ppm (v/v)
Over
range
:
100 ppm (v/v)
Resolution
(ppm)
:
0.1 ppm (v/v)
Detection
Principle
:
Sealed Electrochemical Sensor
Gas
Sampling
:
Diffusion
Accuracy
:
Inherently ±2% (But
also depends on
accuracy of calibration gas)
Drift
:
Less than 2% of signal per month at constant temperature &
pressure
Operating
Life of Sensor
:
Expected life 1 year.
Warranted for six months.
Operating
Temperature
:
0 to 40 °C
Operating
Pressure
:
Ambient ±10%
Maximum
Load resistance
:
500 W
Power
Supply from
Controller
/ DCS
:
16 V – 24 V DC